HS-2 Diffusion Pump

The Agilent HS-2 oil diffusion pump is a pump with high throughput and high pumping speed of 160 L/s air, 200 L/s He, H2. With its fully optimized vapor jet and the robust boiler design, the pump will run without any maintenance for many years. It is available with an ASA inlet flange and with 120 V or 240 V.

This pump is a perfect fit for instrumentation applications such as Gas Chromatography Mass Spectrometry (GCMS) and centrifuges.

  • High pumping performance in a compact package
  • Fractionating jet purifies pumping fluid
  • High tolerable forepressure
  • Stainless steel jet and pump body
  • Standard quick-cool fluid
  • Finned boiler increases surface area, prolonging pumping fluid life
  • Foreline baffle prevents pumping fluid loss into backing system
  • Pumping speed: 160 L/s, maximum throughput: 0.60 Torr L/s
  • Operating range: 3.7 x 10-3 to 5 x 10-8 Torr
  • 15 minute warmup time and 10 minute cooldown time
Part Number Variant

82906301 HS-2 diffusion pump, 2 in ASA, with standard cold cap, 120 V, 450 W
82906302 HS-2 diffusion pump, 2 in ASA, with standard cold cap, 240 V, 450 W
MSPLD10748 HS-2 diffusion pump, ISO 100K, with standard cold cap, 120 V, 450 W
MSPLD10865 HS-2 diffusion pump, ISO 100K, with standard cold cap, 230 V, 450 W

New, factory warranted pumps available.

Price: POA, please contact us for latest prices

Manual

Backstreaming Rate
1 x 10-3 mg/cm2/min
Cooldown Time
10 min
Electrical Requirements
1 ph
50/60 Hz
120/240 V
Fluid Charge
100 CC
Fluid Type
Oil
Input Voltage
120, 240 V
Maximum Forepressure - Full Load
0.40 Torr
0.52 mbar
Maximum Forepressure - No Load
0.55 Torr
0.75 mbar
Maximum Throughput
0.60 T-L/s (0.80 mbar-L/s)
Operating Range
3.7 x 10-3 to <5 x 10-8 Torr
4.9 x 10-3 to <6.5 10-8 mbar
Pump Power
450 W
Pump Weight
10 lbs
4.54 kg
Pumping Speed - Air
160 L/s
Pumping Speed - Air, via AVS 4.1 (1963)
285 L/s
Pumping Speed - He/H2
200 L/s
Recommended Backing Pump
≥ 5.0 cfm (8.5 m3/hr)
Warmup Time
15 min
Water Connections
1/8 inch FPT
Water Cooling - Temperature Requirement
60-80 °F
15-26 °C
Water Cooling Flow Requirement
0.1 gpm
20 L/hr