Precision Vacuum Services Ltd

TEL: 01342 716234
info@precisionvacuumservices.co.uk
Precision Vacuum Services Ltd

NHS-35 Diffusion Pump

The Agilent NHS-35 oil diffusion pump is a pump with high throughput and high pumping speed of 28,000 L/s. With its fully optimized vapor jet, robust boiler design and thermal protection switches, the pump will run without any maintenance for many years. It is available with 35 in. ASA or ISO 1000F (bolted) inlet flange.

This pump is a perfect fit for industrial applications such as vacuum furnace, heat treatment, metallizing, and coating processes for optical lenses, electronics and protective film layers.

  • Five-stage fractionating jet purifies pumping fluid
  • Fill and drain assembly allows for easy exchange of pump fluid
  • Robust boiler design
  • Stable and reliable under widely varying operation conditions
  • High forepressure tolerance
  • Low ultimate pressure
  • High throughput with high pumping speed
  • Fluid level sight glass provides quick indication of pump fluid status at all times
Part Number Variant

MSPLD10786 NHS-35 diffusion pump, ISO 1000F, with standard cold cap, 240 V, 24000 W, with oil temperature sensor 240 V
L5924318 NHS-35 diffusion pump, ISO 1000F, with halo baffle, 240 V, 24000 W 240 V
L5924308 NHS-35 diffusion pump, ISO 1000F, with standard cold cap, 240 V, 24000 W 240 V
F1730308 NHS-35 diffusion pump, 35 in ASA, with standard cold cap, 240 V, 24000 W 240 V
F1730318 NHS-35 diffusion pump, 35 in ASA, with halo baffle, 240 V, 24000 W 240 V
MSPLD10887 NHS-35 diffusion pump, 35 in ASA, with halo baffle, 380 V, 24000 W 380 V
MSPLD10816 NHS-35 diffusion pump, 35 in ASA, with standard cold cap, 380 V, 24000 W 380 V
F1730363 NHS-35 diffusion pump, 35 in ASA, with halo baffle, 400 V, 24000 W 400 V
F1730362 NHS-35 diffusion pump, 35 in ASA, with standard cold cap, 400 V, 24000 W 400 V
L5924336 NHS-35 diffusion pump, ISO 1000F, with halo baffle, 415 V, 24000 W 415 V
L5924326 NHS-35 diffusion pump, ISO 1000F, with standard cold cap, 415 V, 24000 W 415 V
F1730326 NHS-35 diffusion pump, 35 in ASA, with standard cold cap, 415 V, 24000 W 415 V
F1730336 NHS-35 diffusion pump, 35 in ASA, with halo baffle, 415 V, 24000 W 415 V
F1730365 NHS-35 diffusion pump, 35 in ASA, with halo baffle, 440 V, 24000 W 440 V
F1730364 NHS-35 diffusion pump, 35 in ASA, with standard cold cap, 440 V, 24000 W 440 V
L5924364 NHS-35 diffusion pump, ISO 1000F, with standard cold cap, 440 V, 24000 W 440 V
MSPLD10747 NHS-35 diffusion pump, 35 in ASA, with standard cold cap, 460 V, 24000 W, with oil temperature sensor 460 V
L5924319 NHS-35 diffusion pump, ISO 1000F, with halo baffle, 480 V, 24000 W 480 V
L5924309 NHS-35 diffusion pump, ISO 1000F, with standard cold cap, 480 V, 24000 W 480 V
F1730309 NHS-35 diffusion pump, 35 in ASA, with standard cold cap, 480 V, 24000 W 480 V
F1730319 NHS-35 diffusion pump, 35 in ASA, with halo baffle, 480 V, 24000 W 480 V
MSPD10891 NHS-35 diffusion pump, 35 in ASA, with standard cold cap, 575 V, 24000 W 575 V

New, factory warranted pumps available.

Price: POA, please contact us for latest prices

Manual

Backstreaming Rate
5 x 10-4 mg/cm2/min
Cooldown Time
60 min
Electrical Requirements
50/60 Hz
240/415/480 V
Fluid Charge
11356 CC
Fluid Type
Oil
Input Voltage
200, 240, 380, 400, 415, 440, 480, 575 V
Maximum Forepressure - Full Load
0.40 Torr
0.52 mbar
Maximum Forepressure - No Load
0.55 Torr
0.71 mbar
Maximum Throughput
25 T-L/s (33 mbar-L/s)
Operating Range
9 x 10-4 to < 5 x 10-8 Torr
1.2 x 10-3 to < 6.5 x 10-8 mbar
Pump Power
24000 W
Pump Weight
1500 lbs
680.4 kg
Pumping Speed - Air
28000 L/s
Pumping Speed - Air, via AVS 4.1 (1963)
50000 L/s
Pumping Speed - He/H2
35000 L/s
Recommended Backing Pump
≥ 300 cfm (510 m3/hr)
Warmup Time
60 min
Water Connections
3/8 inch FPT
Water Cooling - Temperature Requirement
60-80 °F
15-26 °C
Water Cooling Flow Requirement
4.0 gpm
800 L/hr




Unit 47 Rowfant Business Centre
Wallage Lane
Crawley
West Sussex
RH10 4NQ
Tel: 01342 716234

info@precisionvacuumservices.co.uk



Website By SP Web Connections